A Functional Platform for the Detection of JWH-073 as a Model for Synthetic Cannabinoids
Küçük Resim Yok
Tarih
2018
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Wiley-V C H Verlag Gmbh
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
During the last decade, progress has been made in the on-site detection of abused drug use. Herein, we present an electrochemical biosensor for the detection of one of the synthetic cannabinoids (SCs), JWH-073, using poly (methyl methacrylate) (PMMA) hyperbranched copolymer (HBC) as a base coating and antibody molecules to bind the JWH-073 to the surface. Modification of the surface is proved with various techniques such as differential pulse voltammetry, cyclic voltammetry, and electrochemical impedance spectrometry. The limit of detection and linearity (in logarithmic scale) for JWH-073 were found to be 31.87 ng/mL and 25-500 ng/mL, respectively. The selectivity of the proposed biosensor was also tested in the presence of interfering molecules and the response was much higher for JWH-073 in all measurements. In the final part of the study, synthetic urine was tested with the device and the relative standard deviation value was calculated to be less than 5.0%. The confirmation of data from the resulting bio-platform was performed via LC-QTOF/MS. This work is the first of its kind, a novel, rapid, cost-effective sensing platform for the detection of the SC.
Açıklama
Anahtar Kelimeler
synthetic cannabinoids, surface modification, JWH-073, drug abuse, hyperbranched polymers
Kaynak
Chemelectrochem
WoS Q Değeri
Q2
Scopus Q Değeri
Q1
Cilt
5
Sayı
9