A method to enhance coincidence time resolution with applications for medical imaging systems (TOF/PET)
dc.contributor.author | Ermis, E. E. | |
dc.contributor.author | Celiktas, C. | |
dc.contributor.author | Pilicer, E. | |
dc.date.accessioned | 2019-10-27T22:16:34Z | |
dc.date.available | 2019-10-27T22:16:34Z | |
dc.date.issued | 2014 | |
dc.department | Ege Üniversitesi | en_US |
dc.description.abstract | A coincidence timing spectrometer was assembled using NaI(Tl) inorganic and BC418 type organic (plastic) scintillation detectors. The constant fraction timing method was used. Coincidence time resolution values of the detectors, which have great importance in TOF/PET measurements, were obtained separately. The resolutions were enhanced using a different method in a start-stop coincidence spectrometer: the signals from the start detectors were delayed. The results from the FLUKA Monte Carlo simulation code agreed well with the experimental results. (C) 2014 Elsevier Ltd. All rights reserved. | en_US |
dc.description.sponsorship | TUBITAK, the Scientific and Technological Research Council of TURKEYTurkiye Bilimsel ve Teknolojik Arastirma Kurumu (TUBITAK) [197T087]; EBILTEM, Center of Science and Technology, Ege UniversityEge University [99 BIL 001] | en_US |
dc.description.sponsorship | This work was supported by TUBITAK, the Scientific and Technological Research Council of TURKEY under Project No. 197T087 and by EBILTEM, Center of Science and Technology, Ege University under Project No. 99 BIL 001. | en_US |
dc.identifier.doi | 10.1016/j.radmeas.2014.01.013 | |
dc.identifier.endpage | 59 | en_US |
dc.identifier.issn | 1350-4487 | |
dc.identifier.issn | 1350-4487 | en_US |
dc.identifier.scopusquality | N/A | en_US |
dc.identifier.startpage | 52 | en_US |
dc.identifier.uri | https://doi.org/10.1016/j.radmeas.2014.01.013 | |
dc.identifier.uri | https://hdl.handle.net/11454/50295 | |
dc.identifier.volume | 62 | en_US |
dc.identifier.wos | WOS:000334003200010 | en_US |
dc.identifier.wosquality | Q2 | en_US |
dc.indekslendigikaynak | Web of Science | en_US |
dc.indekslendigikaynak | Scopus | en_US |
dc.language.iso | en | en_US |
dc.publisher | Pergamon-Elsevier Science Ltd | en_US |
dc.relation.ispartof | Radiation Measurements | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | NaI(Tl) inorganic scintillation detector | en_US |
dc.subject | BC418 type plastic scintillation detector | en_US |
dc.subject | FLUKA Monte Carlo simulation program | en_US |
dc.subject | TOF/PET | en_US |
dc.title | A method to enhance coincidence time resolution with applications for medical imaging systems (TOF/PET) | en_US |
dc.type | Article | en_US |