A method to enhance coincidence time resolution with applications for medical imaging systems (TOF/PET)
Küçük Resim Yok
Tarih
2014
Yazarlar
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Pergamon-Elsevier Science Ltd
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
A coincidence timing spectrometer was assembled using NaI(Tl) inorganic and BC418 type organic (plastic) scintillation detectors. The constant fraction timing method was used. Coincidence time resolution values of the detectors, which have great importance in TOF/PET measurements, were obtained separately. The resolutions were enhanced using a different method in a start-stop coincidence spectrometer: the signals from the start detectors were delayed. The results from the FLUKA Monte Carlo simulation code agreed well with the experimental results. (C) 2014 Elsevier Ltd. All rights reserved.
Açıklama
Anahtar Kelimeler
NaI(Tl) inorganic scintillation detector, BC418 type plastic scintillation detector, FLUKA Monte Carlo simulation program, TOF/PET