Development and analysis of modular uniaxial leg adapter

Küçük Resim Yok

Tarih

2007

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Taylor & Francis Ltd

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

The adapters of modular prostheses are the structural components mostly likely to fail during the use of prostheses because of the high forces and moments occurring during the human gait. In this study, structural performance of a modular uniaxial leg adapter during gait is analysed utilizing computer software. By analysing the results, the relevant product was redesigned in order to prevent possible failures. Subsequently, the developed modular uniaxial leg adapter was tested in accordance with ISO 10328 in order to ensure the safety of the design. Further, the test results are compared with a finite element analysis to evaluate the performance of the numerical analysis.

Açıklama

Anahtar Kelimeler

prosthetics, modular prosthetic adapter, structural analysis, FEM

Kaynak

Prosthetics and Orthotics International

WoS Q Değeri

Q4

Scopus Q Değeri

N/A

Cilt

31

Sayı

1

Künye