alpha-Amylase Monitoring by a Novel Amperometric Biosensor Based on Au Electrode: Its Optimization, Characterization, and Application

Küçük Resim Yok

Tarih

2012

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Informa Healthcare

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

A low-cost and sensitive amperometric biosensor was developed for the determination of alpha-amylase activity. The biosensor was constructed by immobilizing glucose oxidase-gelatin via glutaraldehyde on the Au electrode surface. Measurements were carried out chronoamperometrically at -0.7 V. Several parameters such as glucose oxidase activity, gelatin amount, and glutaraldehyde percentage for cross-linking were optimized. Optimum pH, optimum temperature, repeatability, and storage stabilities of the biosensor were identified. Under the optimum experimental conditions, a linear calibration curve was obtained for alpha-amylase between 0.819 and 13.110 U/ml. Sample analyses were carried out by detecting alpha-amylase activities in baker's yeast samples.

Açıklama

Anahtar Kelimeler

alpha-Amylase, Activity determination, Glucose oxidase biosensor, Amperometric biosensor, Au electrode

Kaynak

Artificial Cells Blood Substitutes and Biotechnology

WoS Q Değeri

Q4

Scopus Q Değeri

Cilt

40

Sayı

01.Feb

Künye