Clinical results of implantation of the Medevec® mini-loop haptic anchor plate foldable lenses
dc.contributor.author | Kaskaloglu M. | |
dc.contributor.author | Erakgun T. | |
dc.date.accessioned | 2019-10-27T09:00:38Z | |
dc.date.available | 2019-10-27T09:00:38Z | |
dc.date.issued | 1998 | |
dc.department | Ege Üniversitesi | en_US |
dc.description.abstract | Purpose: To evaluate the surgical performance and postoperative outcomes of 50 consecutive implantations of the Medevec VS 2 UV(TM) mini-loop silicone plate lenses. Setting: Ege University Hospital, Eye Department, Izmir, Turkey. Methods: Forty-three patients of either sex, aged 49 years, or older, were included in the study. Clear corneal incision, circular curvilinear capsulorhexis, phacoemulsification and implantation of the Medevec VS 2 UV(TM) mini-loop silicone plate lenses using the injector-cartridge system was performed in all cases. The mean outcome measures were ease of implantation, intraoperative and postoperative complications, postoperative centration and visual acuity. Results: The mean follow-up was 10.2 months (range 8-11 months). Twenty-two lenses (44 required two or more manoeuvers. In 6 cases, the lens was implanted into a bag with torn anterior capsule and in 3 patients (6 implanted into the sulcus because of a torn posterior capsule during phacoemulsification. Best corrected visual acuity was ? 7/10 in 48 eyes (96%) and ? 10/10 in 42 eyes (84%) at 8 months. The examination of the lenses after 8 months showed no sign of haze, discoloration or tilt. Mean decentration of the lenses was 0.32 mm ± 0.19 (SD). There were no cases of cystoid macular edema or retinal detachment. YAG capsulatomy was performed in one eye. Conclusions: Medevec VS 2 UV(TM) mini-loop silicone plate lenses appeared to have some learning curve for the implantation. They provided - even in the sulcus - excellent centration, good visual acuity and minimal postoperative complications. | en_US |
dc.identifier.doi | 10.1023/A:1006175125099 | en_US |
dc.identifier.endpage | 132 | en_US |
dc.identifier.issn | 0165-5701 | |
dc.identifier.issue | 3 | en_US |
dc.identifier.pmid | 10548456 | en_US |
dc.identifier.scopusquality | N/A | en_US |
dc.identifier.startpage | 129 | en_US |
dc.identifier.uri | https://doi.org/10.1023/A:1006175125099 | |
dc.identifier.uri | https://hdl.handle.net/11454/28098 | |
dc.identifier.volume | 22 | en_US |
dc.indekslendigikaynak | Scopus | en_US |
dc.indekslendigikaynak | PubMed | en_US |
dc.language.iso | en | en_US |
dc.publisher | Springer Netherlands | en_US |
dc.relation.ispartof | International Ophthalmology | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Foldable lenses | en_US |
dc.subject | Phacoemulsification | en_US |
dc.subject | Silicone plate lenses | en_US |
dc.title | Clinical results of implantation of the Medevec® mini-loop haptic anchor plate foldable lenses | en_US |