Reply to Comments on 'Effect of heating rate on kinetic parameters of beta-irradiated Li2B4O7:Cu,Ag,P in TSL measurements'
Küçük Resim Yok
Tarih
2009
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Iop Publishing Ltd
Erişim Hakkı
info:eu-repo/semantics/openAccess
Özet
We appreciate the opportunity to respond to comments regarding the paper published by Ege et al (2007 Effect of heating rate on kinetic parameters of beta-irradiated Li2B4O7: Cu, Ag, P in TSL measurements Meas. Sci. Technol. 18 889). We would like to thank the authors for taking the time to tell us about their opinion, but unfortunately we do not agree with them completely. In the article presented by Kumar and Chourasiya some comment is advanced to the analysis of the glow curves measured with different heating rates, presented in our recent study. According to our study, the area under the glow curve decreases with increasing heating rate in TL-temperature plots due to the quenching effects. Contrary to this, Kumar and Chourasiya suggest that this decrease is due to the normalization process. Here we hope to clarify any confusion regarding our published study.