Electrochemical Determination of Homocysteine at Disposable Graphite Electrodes

Küçük Resim Yok

Tarih

2014

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Wiley-V C H Verlag Gmbh

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

In the case of disruption of Hcy metabolism, the blood level of Hcy increases and it causes particularly the cardiovascular diseases, cancer, dementia and Parkinson's disease. Thus, the sensitive analysis of Hcy levels in biological fluids is very important. Hcy analysis was performed herein using very practical and cost-effective protocol using differential pulse voltammetry and graphite electrode. Detection limit of Hcy was found to be 1.21 mu M in the linear range from 2 mu M to 20 mu M. The electrochemical Hcy detection in artificial urine medium was also successfully performed even in the presence of L-Cysteine, L-Methionine and Glutathione.

Açıklama

Anahtar Kelimeler

Homocysteine, Direct electrochemical detection, Protein, Urine, Pencil graphite electrode

Kaynak

Electroanalysis

WoS Q Değeri

Q2

Scopus Q Değeri

Q2

Cilt

26

Sayı

9

Künye