Standardized method comparison for ACS : 180 plus and Immulite sensitive PSA (sPSA) measurement methods
Tarih
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Erişim Hakkı
Özet
The aim of this study was to compare two immunoassay measurement methods, the ACS: 180 Plus (Chiron Diagnostics) kit and the Immulite sPSA (DPC) kit. Method comparison analysis was performed according to EP9-A; approved guideline of NCCLS 51. Serum samples having a wide range of total prostate-specific antigen (PSA) concentrations were evaluated in split-sample analysis. F-test, t-test analysis and regression statistics were performed. In Deming regression analysis the coefficients were as follows; the slope=0.967; y-intercept=-0.148, r=0.989. An acceptable bias was seen since the systematical error was calculated to have a value less than the total allowable error calculated from biological variations. Nonparametric evaluation of the area under ROC curves for ACS: 180 Plus and Immulite sPSA were 0.997 and 0.987, respectively. Diagnostic accuracy was at the level of p=0.000 and no statistical difference was found between the two assay methods.